Ascott CC450iP Cyclic Corrosion Test Chamber
450 Ltr (15.8 cu ft)
This instrument purchased new in March 2018 and was only used on one occasion hence is in exceptionally clean condition and in fact can be regarded "as new"
Low loading threshold for ergonomic loading and unloading.
‘Easy Open’ pneumatically operated canopy, with safety interlock.
Dry seal gasket to prevent wetting the operator’s clothes.
Timed canopy lock and purge of the cabinet interior with fresh air after testing and before the chamber is opened.
Easily accessible salt spray atomizer(s) in transparent hard-wearing acrylic, for ease of maintenance.
External 80L portable salt solution tank for ease of cleaning.
Central viewing window for unimpeded visual access to chamber interior and test material.
Base castors and fork integrated lift lift truck runners, for ease of transportation.
Intuitive touch-screen, full color user control interface, employing the highly intuitive operating software, for ease of programming and use.
The operating software has a very large capacity for the creation of complex multi-step test programmes, so that the widest possible range of test profiles can be accommodated.
The icon based control interface offers a variety of languages built-in for ease of understanding and use.
RJ45 communications port enabling the chamber to be wired or wirelessly connected to a local area network (LAN) for logging and remote programming via a computer running optional software.
A real-time, user configurable ‘clock’ to enable different test samples to be batched and have their exposure time monitored separately, with alarms to warn when a pre-set test period has been completed.
‘Easy Open’ pneumatically operated canopy, with safety interlock.
Dry seal gasket to prevent wetting the operator’s clothes.
Timed canopy lock and purge of the cabinet interior with fresh air after testing and before the chamber is opened.
Easily accessible salt spray atomizer(s) in transparent hard-wearing acrylic, for ease of maintenance.
External 80L portable salt solution tank for ease of cleaning.
Central viewing window for unimpeded visual access to chamber interior and test material.
Base castors and fork integrated lift lift truck runners, for ease of transportation.
Intuitive touch-screen, full color user control interface, employing the highly intuitive operating software, for ease of programming and use.
The operating software has a very large capacity for the creation of complex multi-step test programmes, so that the widest possible range of test profiles can be accommodated.
The icon based control interface offers a variety of languages built-in for ease of understanding and use.
RJ45 communications port enabling the chamber to be wired or wirelessly connected to a local area network (LAN) for logging and remote programming via a computer running optional software.
A real-time, user configurable ‘clock’ to enable different test samples to be batched and have their exposure time monitored separately, with alarms to warn when a pre-set test period has been completed.
Wetting mode
Temperature range
Adjustable from ambient to +70°C
Humidity range Fixed at 95% – 100% RH
Temperature range
Adjustable from ambient to +70°C
Humidity range Fixed at 95% – 100% RH
Salt spray mode
Temperature range
Adjustable from ambient to +50°C (This instruemnt fitted with ACC03/1 Extras Heaters to achieve 60°C during Salt Spray)
Salt spray fall-out rates
Adjustable from 0.5 to 2.5 ml per 80 cm² per hour
Temperature range
Adjustable from ambient to +50°C (This instruemnt fitted with ACC03/1 Extras Heaters to achieve 60°C during Salt Spray)
Salt spray fall-out rates
Adjustable from 0.5 to 2.5 ml per 80 cm² per hour
Drying mode
Temperature range Adjustable from ambient to +70°C
Temperature range Adjustable from ambient to +70°C
Chamber construction
Glass reinforced plastic, Polypropylene & PVC parts
Glass reinforced plastic, Polypropylene & PVC parts
Electricity supply 3 phase Voltage (VAC) 380 - 440V (50Hz)
Water - Deionized/distilled for topping up air saturator and making salt solution. Air saturator requires a continuous water connection 1.5-5.0 bar (21-72 psi). If ACC42 is ordered the continuous water supply must be pressuried 4.0 – 5.0 bar (58-72 psi). If air saturator is topped up manually option ACC66 must be ordered.
Air
Clean dry & oil free, 4.0 to 6.0 bar (58-87psi) with 240 Ltrs (8.5cu.ft) per minute flow
Clean dry & oil free, 4.0 to 6.0 bar (58-87psi) with 240 Ltrs (8.5cu.ft) per minute flow
Exhaust
Exhaust – 3m (10ft) exhaust pipe is required which should be terminated outside building (see accessory ACC58)
Exhaust – 3m (10ft) exhaust pipe is required which should be terminated outside building (see accessory ACC58)
Drain
3m (10ft) drain pipe is required which should be terminated into floor level drain (see accessory ACC58)
3m (10ft) drain pipe is required which should be terminated into floor level drain (see accessory ACC58)
Operating environment conditions
+18 to +23°C (+64 to 73°F), 85% max RH (non condensing) ambient.
+18 to +23°C (+64 to 73°F), 85% max RH (non condensing) ambient.
Ext dims (D) 64cm x (W) 101cm x (H) 114cm
ACC120 Logging Software on USB stick
SKU | P006 |
---|---|
Availability | Out of Stock |
Condition | Used |
Shipping | £60.00 |